Products > Insufflators > CO2 Change Over Valve

Technical Data  
Type CO2 Change Over Valve
Proof Pressure 0 - 200 bar
Gas Carbon Dioxide Medical Grade
Gas Connectors Optional: Pin-Index, UNF, German standard
Temperature Range 10° to 50 °C ; 50° to 104°F
Weight 1,2 kg
Dimensions 155 x 170 x 120 mm




Download Adobe Acrobat Reader
PDF 498 kb | PDF 213 kb Adobe Acrobat Reader
 
CO2 Change Over Valve
Do you know this problem?
In the midst of an operation the gas flow from your Insufflator stops!

We have a solution! An automatic mechanicals
CHANGE-OVER VALVE


This automatic change-over valve makes continuous gas insufflation possible and results in an uninterrupted operation.

Q. "Why does this problem exist at all, even though the insufflator is equipped with a gas consumption indicator?


A. "The problem is caused by a relatively unknown feature of carbon dioxide. Contrary to other gases, such as oxygen, carbon dioxide when stored under pressure - such as in a gas bottle - exists in liquid state. This causes a constant bottle pressure as long as there is some liquid carbon dioxide in the bottle. In other words, even a nearly empty bottle appears to have maximum pressure. Only if the liquid gas is completely consumed, will the bottle pressure decrease.

Diagram 1: Gas Bottle Pressure in Relation to Amount Used
Diagram 1: Gas Bottle Pressure in Relation to Amount Used

As you can see the meter indicates a decrease in bottle pressure only when the liquid gas is used up - then it is almost too late. In contrast with meters for other types of gases or the petrol gauge in your car you can see more precisely how much is left in the tank.

During an operation, if the first bottle of gas is used up, changing over to a second bottle will be carried out automatically. This bottle will be completely emptied except for a small residue (below 1 %) to prevent contamination of the bottle. After the operation the empty bottle can simply be replaced.





To reliably carry out an operation bottles are often replaced which are not completely empty. To eliminate problems with limited gas supply and the unnecessary replacement of bottles we developed the new automatic mechanical gas bottle change-over valve. Due to its pure mechanical construction it can easily fit into your existing operating system."

 


-------------------